SC 340 Solar Simulation Chamber (MH/MHG)
Full function climatic chamber with additional simulation of global sunlight spectra. With its 340 l test chamber, the SC340 is suited to test material samples and component sections.
- Radiation source: Metal Halide Lamp, 1 x 1500 W (Type MH); 1 x 1200W (Type MHG)
- Chamber volume: 580 x 765 x 750 mm (WxDxH)
- Filter system: Outdoor/Indoor
- Irradiance: 1000W/m2 (Type MH) ; 800-1200 W/m2 (Type MHG)
- Spectral distribution: Global Radiation 280 ? 3000 nm (Base: CIE, Table 4; Share: DIN 75220, Table 1, Column 2/4)
- Temperature ranges w/ radiation: -20 °C to 100 °C
- Temperature ranges w/o radiation ?30 °C to 100 °C
- Heating rate: 3.0 K/min
- Cooling rate: 2.5 K/min
- Irradiated area approx. 1600 cm²
- Touch control panel
- Micro-processor monitoring and control unit SIMCON/32
- Digital customer I/O, potential-free 24 V, 4 freely disposable inputs/outputs
- Independent adjustable temperature limiter tmin / tmax
- Adjustable software temperature limiter min./max.
- Humidity input and display in % rel. humidity
- Centronics printer interface
- Serial interface RS 232
- Potential-free contact for switching-off of test specimens
- European socket
- Calibration of 2 temperature values and 2 climate values
- Air-cooled refrigeration unit
- Condensation protection (dehumidification during heating phase)
- Aerosol-free humidification and dehumidification
- Psychrometric humidity measuring system
- Water supply tank for humidification water with level indication
- Automatic water replenishment with low water alarm
- 1 Entry port, NW 50 mm, left-hand side
1 Entry port, NW 125 mm, right-hand side
- 1 Shelf made of stainless steel
- 1 Set of operating instructions
- Temperature range +15 °C to +80 °C
Temperature fluctuation
with steady state conditions
according to IEC 60068-3-5
depending on the test space
temperature temporal d±0.5 K
Humidity range 10 % to 80 % r. h.
Dew point range +5 °C to +74 °C
Humidity fluctuation
with steady state conditions
depending on climatic value temporal ±3.0 % to ± 5.0 r. h.
- Temperature range approx. -20 °C to +100 °C
Temperature fluctuation
with steady state conditions
according to IEC 60068-3-5
depending on the test space
temperature temporal d±1.0 K
- Radiation intensity 1000 W/m²
in the range from 280 to 3000 nm,
in a distance of 600 mm
Homogenity ±5 %
Spectral radiation performance according to DIN 75220 Tab 1 print 2/4
Kind of radiation metal-halogenid lamp, 1 x 1,5 kW
Irradiated area approx. 1600 cm²
- Temperature range approx. -30 °C to +100 °C
approx. -40 °C to +120 °C (light transmission
aparture closed with optional isolation cover)
Temperature fluctuation
with steady state conditions
according to IEC 60068-3-5
depending on the test space
temperature temporal ±0.1 K to ±0.5 K
Temperature rates of change
according to IEC 60068-3-5 heating: approx. 3.0 K/min.
cooling: approx. 2.5 K/min.
Calibrated temperature values +23 °C and +80 °C
- Temperature range +10 °C to +90 °C
Temperature fluctuation
with steady state conditions
according to IEC 60068-3-5
depending on the test space
temperature temporal ±0.1 K to ±0.3 K
Humidity range 10 % to 90 % r. h.
Dew point range +5 °C to +87 °C
Humidity fluctuation
with steady state conditions
depending on climatic value temporal d±3.0 % r. h.
Calibrated climatic values +23 °C / 50% r.h. and +90 °C / 50% r.h.
- Insulated cover: Expands the temperature range to -40°C to +120°C
- Misting equipment: Expands the test possibilities in the area of climate
- UV sensors: 280 nm to 320 nm / 320 nm to 400 nm
- Control of black standard temperature
- Black standard temperature sensor
- SolarTest sensor: 300 - 3000 nm
- Printer: 6-color, programmable
- S!MPATI software for Data Acquisition
- RS 422/485 interface
Reference Test Specifications include:
Reference Test Specifications include:
- EPA or others.
-
DIN 75220
Ageing of automotive componentsin solar simulation units
-
MIL-STD 810
Environmental engineering considerations and laboratory tests